Technical Specifications

Element rangeChlorine (Cl, 17) to Uranium (U, 92) (up to 55 elements simultaneously)
DesignFully enclosed X-Ray system with mechanical positioning mechanism for up to 4 measurement positions

 

Dimensions

External Dimension:

W x H x T

Instrument only : 400 x 490 x  170 mm (15.7 x 19.3 x 6.7 in)

Instrument + Enclosure : 800 x 1000 x 300 (31.5 x39.4 x 11.8

Weight14 kg (31 lb.) + 39 Kg ( 86 lb. ) enclosure

 

X-ray Source and Detection

X-ray tubeLow Power X-ray tube 4W with Rh or W target
High voltageThree settings: 30 kV; 40 kV; 50 kV
Primary filterChangeable:  Al 250 µm; Al 500 µm Al 750 µm
Collimator1 or 2 mm Ø
X-ray detectorSilicon-drift-detector (SDD)
Detector area20 mm²
Detector resolutionFwhm < 130 eV (at Mn-Kα)

 

Electrical and Environmental data

Power supplyAC 100-240 V    50 -60 Hz
Power consumption≤ 70 VA

 

Environmental conditions

Operating temperature10°C – 50°C (50°F – 104°F)
Storage temperature0°C – 50°C (32°F – 122°F)
Humidity≤ 95 %, non-condensing

 

Standards

CE approvalEN 61010 , EN 61326
X-ray standardsDIN EN ISO 3497, ASTM B 568
ApprovalIndividual acceptance approval per §12 StrlSchG   (German radiation ordinance) Complies to US Standard ANSI/HPS N43.3 Radiation Safety for X-ray diffraction and fluorescence analysis equipment